The CELL-tainer bioreactor system provides accurate, low shear mixing and high mass transfer rates. Mixing of cell cultures is essential to create and maintain homogenous circumstances in the bioreactor. The single-use bioreactor bag is positioned in the tray and placed on the rocking platform creating a two-dimensional movement (including a wave like motion) in the culture medium. The rocker angle and frequency can be adjusted to increase or decrease the mixing time and the oxygen transfer rate.
The CELL-tainer control system has three password protected access levels. The access to the local human interface can be blocked by the supervisory software allowing the system to be used in compliance with 21 CFR part 11.
The robust design and heavy-duty motor guarantee years of trouble-free operation at extreme low noise levels.

ParameterSpecifications
Dimensions unitW x D x H = 1400 x 805 x 565 mm (approx.)
Dimensions platformW x D x H = 660 x 510 x 100
Removable tray (bag holder)
Light protectionClosed cabinet with frontloading (drawer). The front door is provided with a window which is UV protective and lux reducing. The interior of the cabinet can be made visible by switching (pushbutton) a non-UV LED light.
Movement profile2D movement: horizontal displacement combined with vertical displacement. Tilt is manually adjustable.
 Tilt angle: 5o – 17o; resolution 1.7o; accuracy 1.7o
Default angle 8.5o
 Rotational speed 5- 20 rpm (default); resolution 1 rpm; accuracy 0.5 rpm
Temperature controlSensor on platform, directly under the bag; accuracy ± 0.2oC
Sensor in incubator room; accuracy ± 0.3oC
Control range: 20 - 40oC (ambient: 15 - 25oC)
Resolution: 0.5oC
Heating capacity2 x heating ventilator (230 W each)
DO measurementDissolved Oxygen amplifier with disposable polarographic sensor fixed mounted in the bottom of the bag.
 5 – 150 % air saturation (+/- 5% air saturation) Air flow control by direct reading flowmeter 16 Ln/min
DO control (PID)Direct reading flowmeter / solenoid valve combination: 0 – 1Ln/min O2
pH measurementpH- amplifier with disposable electrochemical sensor fixed mounted in the bottom of the bag.
 pH range: 2.00 – 10.00 (+/- 0.03 unit)
pH control (PID)
Alkaline:peristaltic pump Watson Marlow 313, 24VDC, 45 rpm.
 Pumphead 313:  applicable with 1.6mm wall tubing.
  Complete with adjustable tube clamps
  for 0.5 mm to 8.0 mm ID tubing
 
Acid:Flowmeter / solenoid valve combination for CO2: 0 – 1Ln/min
Air supply1-3 barg.
 Additional gasses (O2, CO2): 3 barg Gasses internally reduced to 0.1-0.2 barg max.
 Maximum bag pressure (operating) 0.2 barg
Exhaust gasEasy-Load filter heater for exhaust gas filter. Range: 5-50oC; setpoint 10oC above setpoint bioreactor. Cut-off at 50oC
Process monitoring and controlSingle Board Computer & Display unit
 SBC
CPU ARM9 300MHz
32MB Flash / 32MB RAM
3 x RS232 Serial interfaces
1 x Ethernet UTP interface
1 x SD-card slot
Optional CAN 2.0 interface (DeviceNet)‏
Embedded Linux OS
 DISPLAY
5.7 inch color CSTN graphical LCD
320x 240 pixels
4-wire touchscreen
Power
12Vdc 200mA
Dimension
168 x 109 x 38 mm
 I/O Module    MCU
MCU Microchip PIC24FJ256GB106 (32MHz)
256KB Flash / 16KB RAM
1 x RS232 Serial interface
4 x 1-Wire interface for DS18S20 (temperature sensor)
12 x Optical isolated digital inputs
18 x Relais outputs
4 x Analog inputs
Power
24Vdc 300mA
Dimension
220 x 130 x 55 mm
Temperature sensor
DS18S20; accuracy 0.25°C using extended resolution
Data storage and operation- Measuring rpm, T, DO, pH up to 1/s (adjustable sampling time)
- Data storage / batch
- Alarm logs: door open, change set-points, not reaching set-point
- Dose monitoring: show total sec/ml or percent in the
   Process View screen
- Manual operation of valves and pumps
- Manual position of tray (jog, sample, drain)
Data exportEthernet UTP interface (xml.file) to be approached from web browser (via USB)
SafetyDoor opening / safety switch (stop) on all functions
Protection CE/IP54
Mains230 Vac / 50 Hz – 115 Vac / 60 Hz
Cable cord & plugTo be specified with order
ManualsOperation & Installation


Skid CMS2010221
The open frame of aluminum is completed with blockable wheels (4x) and adjustable feet (4) to fixate the skid on the position of use. Optional (additional price)are drawers which can be used to hold medium bags or others.


Footprint:
Material frame:
Material plating, covering:
Wheels:
Floor clearance:
1400 x 830mm
blanc anodized aluminium (80x80mm)
Trespa, off-white
GMP qualified when moving
200mm